Powered by ZigaForm version 3.9.9.7
Hot

Kaye Validator AVS

Advanced Validation Technology

The Kaye Validator AVS (Advanced Validation System) combines accurate sensor measurements with all GMP requirements for calibration and traceability to national standards, while generating compliant reports and managing the validated assets and validation equipment. The Validator AVS offers improved hardware features, an easy to use data handling concept and an intuitive, user-friendly software interface.

Description

Reducing Your Cost of Ownership

With our integrated validation system and improved data management concept, the Kaye Validator AVS saves time and money by simplifying the entire validation process.

Lifting Validation to the Next Level

The Kaye Validator AVS introduces a completely new validation concept. The AVS console will be the common platform for all validation products. It is dedicated for validation tasks only – so no worries about operating systems. The Software is preloaded – with a customized interface for validation equipment, simplifying the validation process.

FEATURES

Validator AVS

  • Enhanced data handling and redundancy (4-levels)
  • Expanded input capacity, types, and range
  • Improved scan time 2/sec for 48 inputs (EN554/ISO17665)
  • Stand-alone operation with Console docking station

Validator AVS Console

  • Hardened, rugged Console pre-loaded with all Kaye software
  • Easy-to-operate / state-of-the art intuitive user interface
  • Ethernet/ WiFi/ Docking Station connections
  • Asset centric concept for process equipment and Kaye hardware
  • Console capable of running existing Kaye product software.
  • Improved Reporting Analysis Tool –backward compatible
  • Simplified validation (GAMP-5)

Applications

  • Steam Sterilizers (Autoclaves)
  • Dry Heat Sterilizers
  • Steam in Place (SIP)
  • Water Cascade/Fall Sterilizers
  • Incubators
  • Stability Chambers
  • Freezers
  • Freeze Dryer/Lyophilisation
  • Vessels

Technical specifications

Analog InputUp to 48
ThermocouplesType T, J, K,E,B,R,N,S: 0.1°C; T+ limited range 0.01°C resolution
Scanning Speed36 channel in 1 sec and 48 channels in 2 Sec
Internal Memory4 gb for data collection
Input Impedance10KΩ. Source greater than 10KΩ produces open circuit indication
Common Mode Rejection160 db (8 inputs/sec) @ line frequency
145 db (12 inputs/sec) @ line frequency
140 db @ DC
Max. Common Mode Voltage100V pk ch-to-ch350V pk ch-to ch to frame ground
Normal Mode Rejection82 db @ 60 Hz (8 inputs/sec)69 db @ 60 Hz (12 inputs/sec)
Voltage Input0 to 10 VDC
Resolution1:72,000
Voltage Input Accuracy30 days: ±(0.003% of reading + 2 counts + 4 microvolts)
1 year: ±(0.006% of reading + 2 counts + 4 microvolts)
Sensitivity0.5 microvolts/count on most sensitive range
Voltage Temp. Coef.±(0.1 microvolts + 0.001% reading)/°C
Compensator Temp. Coef.±0.01°C per °C
Input Terminal Temperature
Non-uniformity
±0.1°C from calibrated terminal
Input Ranges-6 to 30mV, -12 to 60mV, -60 to 300mV, -2 to 10V
EnvironmentalTemperature: 0 to 50°C (32 to 122°F)
Relative humidity: 95% non-condensing
Power90 to 250 VAC, 50/60 Hz
Fuse Rating4A Slow Blow
Size190H X 411W X 381 mm D (457 mm with SIM)
7.5 in H x 16.2 in W x 15 in D (18 in with SIM)
Weight10.60 kg (23.4 lbs)
BatteryLithium ion with minimum 3 hours of battery backup